Radiation Exposure Converter
Convert radiation exposure between units in real time. Enter a value, choose the from and to units, and read the result instantly.
What radiation exposure measures
Exposure quantifies the ionisation that X-rays or gamma rays produce in air. It is the charge liberated per unit mass, measured in coulombs per kilogram (C/kg), with the roentgen (R) still used on survey meters and calibration certificates. One roentgen is exactly 2.58 × 10⁻⁴ C/kg.
The quantity is defined for photons in air only. It says nothing about what happens inside tissue, and it cannot be applied to beta particles, neutrons or alpha emitters. That restriction is the single most important thing to know before you convert a reading.
The conversion rule
The roentgen is fixed by definition at 2.58 × 10⁻⁴ C/kg, so both R factors are exact rather than measured. Milli adds another three zeros, which puts 1 mR at 2.58 × 10⁻⁷ C/kg.
The milli and roentgen pairs interact, so a 100 mR reading is 100 × 2.58 × 10⁻⁷ = 2.58 × 10⁻⁵ C/kg, which is also 0.0258 mC/kg. Two conversions in one figure, and both are simple decimal shifts.
Worked example: 100 mR
The defaults convert 100 mR to C/kg. Multiply by 2.58 × 10⁻⁷: 2.58 × 10⁻⁵ C/kg. In millicoulombs the same value is 0.0258 mC/kg, and the four-decimal display rounds the SI figure to 0.0000.
That reading is an older entrance exposure for a chest radiograph. The air kerma at that surface is about 0.9 mGy, since the chain from exposure to energy deposition runs at 8.77 mGy per roentgen.
Conversion table
| mR | R | C/kg | Air kerma |
|---|---|---|---|
| 1 | 0.001 | 2.58 × 10⁻⁷ | 8.77 µGy |
| 10 | 0.01 | 2.58 × 10⁻⁴ | 87.7 µGy |
| 100 | 0.1 | 2.58 × 10⁻⁵ | 0.877 mGy |
| 1,000 | 1 | 2.58 × 10⁻⁴ | 8.77 mGy |
| 5,000 | 5 | 1.29 × 10⁻ | 43.9 mGy |
The 1 mR row is the scale of scatter behind a protective screen, 100 mR matches an older chest-image entrance exposure, and 5,000 mR is 5 R, a level reached in prolonged fluoroscopy and tracked as a dose alert. Each step is a factor of ten in exposure and a factor of ten in air kerma.
Where exposure readings are used
Exposure survives because ion chambers are calibrated in it, and because it is the quantity a survey meter can measure directly without knowing the beam energy in detail.
- X-ray room surveys, where leakage and scatter are logged in mR/h.
- Ion chamber calibration against a free-air standard at a national laboratory.
- Dental and mammography quality assurance, where entrance exposure is checked per unit.
- Historical film badge records, which report cumulative exposure in roentgens.
How a chamber turns exposure into charge
A 100 cm³ air chamber holds 1.2 × 10⁻⁴ kg of air at room density. Exposed to 100 mR, it collects 2.58 × 10⁻⁵ × 1.2 × 10⁻⁴ = 3.1 nC. Electrometer design is built around charge that small, which is why cable leakage and humidity dominate the error budget.
Exposure versus absorbed dose
Exposure measures air ionisation; absorbed dose measures energy deposited per unit mass in any material, in gray. The bridge is the mean energy per ion pair, 33.97 J/C, so 1 R corresponds to 8.77 mGy in air. Tissue converts with a similar factor, near 9.5 mGy per roentgen, depending on beam quality.
Equivalent dose in sievert adds a radiation weighting factor, which is 1 for X-rays and gamma rays but 20 for alpha particles. A 1 mGy whole-body dose of gamma radiation is therefore 1 mSv, while the same absorbed dose from alpha emitters would be 20 mSv. Exposure in C/kg does not see that difference at all.
Dose limits expressed as exposure
Radiation protection limits are written in sieverts, so converting them to exposure needs the 8.8 mSv per roentgen bridge for gamma radiation. The ICRP occupational limit of 20 mSv a year corresponds to about 2.3 R of whole-body gamma exposure, and the 1 mSv public limit to roughly 0.11 R.
Those figures are only a sanity check, not a compliance method. A real assessment accounts for beam energy, geometry and tissue weighting, and an occupancy factor for the time actually spent in the field. Treat the roentgen equivalent as an order-of-magnitude cross-check against a survey meter reading.
Precision and common mistakes
- Applying roentgens to beta or neutron fields, where the unit is undefined.
- Treating 100 mR as 100 mSv; the two are different quantities and the numeric coincidence is accidental.
- Reading a rate in mR/h as an integrated exposure without multiplying by time.
- Forgetting that exposure depends on photon energy, so a 20 keV and a 1 MeV beam of equal C/kg deliver different doses.
Unsealed chambers need an air-density correction for temperature and pressure. A 3 percent rise in barometric pressure puts 3 percent more air mass in the same volume, so the same photon field liberates 3 percent more charge and the meter over-reads by that amount until the correction is applied. Calibration certificates state the reference conditions, usually 20 °C and 101.3 kPa.
Rates and integration
Survey meters usually read mR/h, while records are kept in mR. A room reading 2 mR/h occupied for 30 minutes accumulates 1 mR, which is 2.58 × 10⁻7 C/kg. Doubling the occupancy doubles the accumulated exposure, so the integration is linear and no decay correction applies to the instrument reading itself.
Frequently asked questions
How many C/kg is 1 R?
Exactly 2.58 × 10⁻⁴ C/kg. The roentgen has been defined that way since 1928, so the factor is a definition rather than a measurement.
How many mR is 100 mC/kg?
387,600 mR, which is 387.6 R. Divide 0.1 C/kg by 2.58 × 10⁻⁴ to get roentgens, then multiply by 1,000.
What is 100 mR in C/kg?
2.58 × 10⁻⁵ C/kg, or 0.0258 mC/kg. The display shows 0.0000 because the value sits five decades below one.
Is a roentgen the same as a rad?
No. The roentgen measures air ionisation, the rad measures absorbed energy. One roentgen corresponds to about 0.877 rad in air, which is 8.77 mGy.
How many mGy is 1 R in tissue?
About 9.5 mGy for diagnostic beams, against 8.77 mGy in air. The difference comes from the energy absorption coefficients of tissue and air.
Can I use C/kg for beta radiation?
No. Exposure is defined only for photons in air. For beta emitters, measure absorbed dose in gray with a suitable detector instead.
What exposure does a chest X-ray give?
Roughly 50–100 mR at the entrance surface, which is 1.3–2.6 × 10⁻⁵ C/kg and about 0.4–0.9 mGy of air kerma. Digital systems with tighter collimation often run at the low end.
How do I convert mR/h to a total exposure?
Multiply by the hours of occupancy. At 2 mR/h for 0.5 h you accumulate 1 mR, which is 2.58 × 10⁻7 C/kg.
Why is 1 mR not 1 mSv?
Because exposure and equivalent dose measure different things. For gamma rays, 1 R in air is about 8.8 mGy, and with a weighting factor of 1 that is 8.8 mSv, not 1,000 mSv.
Method and sources
- Linear conversion through C/kg: result = value × factor(from) ÷ factor(to).
- Factors applied: 1 mC/kg = 0.001 C/kg; 1 Roentgen (R) = 0.000258 C/kg; 1 Milliroentgen (mR) = 2.58 × 10⁻⁷ C/kg.
- The millicoulomb prefix is exact, and the roentgen is defined as exactly 2.58 × 10⁻⁴ C/kg of air, so the milliroentgen is exactly 2.58 × 10⁻⁷ C/kg.
- Every factor is an exact definition or an exact decimal SI prefix, so the only error is the floating-point rounding of the arithmetic itself.