slug: solution-dilution-calculator
title: "Solution Dilution Calculator"
category: scientific
updated: 2026-11-17
canonical: https://calcseed.com/c/scientific/solution-dilution-calculator/
html: /c/scientific/solution-dilution-calculator/
---

# Solution Dilution Calculator

Solve the dilution equation C₁V₁ = C₂V₂ to find stock volume, final volume or concentration. Free and instant.

## Quick answer

**Quick answer:** with the default inputs (1 M, 0.1 M, 1,000 mL), the stock volume is **100 mL**. Change any input to update the result.

## Default calculation

Values computed by this calculator at exactly the default inputs listed below.

### Inputs (defaults)

| Input | Value | Unit |
| --- | --- | --- |
| Stock concentration | 1 | M |
| Final concentration | 0.1 | M |
| Final volume | 1000 | mL |

### Outputs at these defaults

| Output | Value | Unit |
| --- | --- | --- |
| Stock volume | 100 | mL |

## Formula

```text
C₁V₁ = C₂V₂
```

## Inputs

| Variable | Type | Unit | Default |
| --- | --- | --- | --- |
| Stock concentration | number | M | 1 |
| Stock volume | number | mL | 100 |
| Final concentration | number | M | 0.1 |
| Final volume | number | mL | 1000 |

## Outputs

| Result | Format |
| --- | --- |
| Stock volume | primary |

## How it works

### What the dilution equation assumes

Adding solvent changes the concentration but not the amount of solute. That single conservation statement is the whole equation: moles before the dilution equal moles after it, and since moles = concentration × volume, C&sub1;V&sub1; = C&sub2;V&sub2;. This calculator solves for whichever of the four values is unknown.

The volume on the left is the volume of stock you measure out, not the volume of solvent you add. V&sub2; is the final total volume after topping up, so the solvent you actually pour is V&sub2; − V&sub1;. Confusing those two is the most common arithmetic slip in the lab.

Units only need to be consistent within each pair. Molarity with millilitres works as well as molarity with litres, because the volume units cancel on both sides. The concentration units must match each other, and the volume units must match each other, but the two pairs are independent.

### Worked example at the defaults

The defaults describe making 1 L of 0.1 M solution from a 1 M stock. The stock needed is V&sub1; = C&sub2;V&sub2; ÷ C&sub1; = 0.1 × 1,000 ÷ 1 = **100 mL**, which is what the calculator returns.

In mole terms, 100 mL of 1 M stock contains 100 mmol, and diluting that to 1,000 mL gives 100 mmol ÷ 1 L = 0.1 M. The solute never changed; only the volume it sits in grew tenfold, so the concentration fell tenfold.

The practical instruction is: measure 100 mL of stock into a 1,000 mL volumetric flask, then add solvent up to the mark. That means adding about 900 mL, not 1,000 mL. Adding a full litre of water to 100 mL of stock would give 1,100 mL at 0.0909 M, a 9% error that quietly fails a calibration curve.

### Dilution factors at a glance

| Dilution | Stock per 100 mL final | Result from a 1 M stock |
| --- | --- | --- |
| 1:2 | 50 mL | 0.5 M |
| 1:5 | 20 mL | 0.2 M |
| 1:10 | 10 mL | 0.1 M |
| 1:20 | 5 mL | 0.05 M |
| 1:50 | 2 mL | 0.02 M |
| 1:100 | 1 mL | 0.01 M |
| 1:1000 | 0.1 mL | 0.001 M |

A 1:10 dilution means one part stock plus nine parts diluent, giving ten parts total. The colon notation is the reverse of the concentration ratio, which trips people up: 1:10 produces a tenfold reduction, not a tenfold increase. Read it as stock to total whenever the result looks surprising.

### Serial dilution, step by step

Going from 1 M to 0.001 M in one step needs 1 mL of stock in a litre, which is hard to measure accurately. Four tenfold steps use comfortable 10 mL volumes and compound the error only slightly, which is why serial dilution is standard practice for anything below about 0.01 M.

| Step | Stock taken | Final volume | Resulting concentration |
| --- | --- | --- | --- |
| Start | — | 100 mL of 1 M | 1 M |
| 1:10 | 10 mL of 1 M | 100 mL | 0.1 M |
| 1:10 again | 10 mL of 0.1 M | 100 mL | 0.01 M |
| 1:10 again | 10 mL of 0.01 M | 100 mL | 0.001 M |
| 1:10 again | 10 mL of 0.001 M | 100 mL | 0.0001 M |

Each step multiplies the concentration by 0.1, so four steps give a 10,000-fold reduction. The trade-off is error propagation: a 0.5% pipetting error at each of four steps becomes roughly 1% overall, still better than a single 1,000-fold transfer measured to the same tolerance.

### Doing it at the bench without losing accuracy

- Use a volumetric flask for the final volume, not a graduated cylinder; the flask is calibrated to 0.05 mL at 1,000 mL while a cylinder may be off by 5 mL.
- Rinse the pipette with the stock twice before measuring, so the delivered volume is not diluted by residual water.
- Add the concentrated stock to part of the solvent first, then top up, which improves mixing and reduces local heating in acid dilutions.
- Let the flask reach room temperature before adjusting to the mark if the dilution released or absorbed heat.
- Invert the stoppered flask at least ten times; a swirl is not enough to homogenise a dense stock at the bottom.

### Where the equation stops being exact

The equation treats volumes as additive, which is close for dilute aqueous solutions and wrong for concentrated ones. Mix 50 mL of ethanol with 50 mL of water and you get about 96.4 mL, not 100 mL, because the smaller water molecules pack into the gaps between the larger ethanol molecules. A nominal 1:1 dilution therefore lands 3.7% above the predicted concentration.

The same contraction shows up with strong acids and bases at high concentration. Diluting 100 mL of concentrated sulfuric acid to 1 L does not give a solution whose volume is exactly the sum of the parts, and the heat released changes the density besides. For anything above about 0.5 M, prepare by weight and convert with the measured density instead of trusting the volume.

#### When the solute reacts or evaporates

If the solute dissociates into several ions, the equation still holds for the formal concentration but not for each species: 0.1 M calcium chloride is 0.1 M in CaCl_2_ units and 0.2 M in chloride ions. If the solute is volatile, such as ammonia or acetic acid, the concentration of a stored dilute solution drifts downward as the solute leaves, and C&sub1; is no longer what the label says.

#### Volumetric glassware has a temperature

A flask marked 1,000 mL is calibrated at 20 &deg;C. Fill it at 5 &deg;C and the glass and the water both contract, and the delivered volume is about 0.1% short; at 35 &deg;C it is 0.1% long. That is small for most work and not small for a 0.1% assay, which is why certified methods specify equilibration to room temperature.

## FAQ

### How do I make a dilution from a stock solution?

Solve V&sub1; = C&sub2;V&sub2; ÷ C&sub1;, measure that volume of stock, then add solvent to reach the final volume. For 1 L of 0.1 M from 1 M stock, measure 100 mL and top up to 1 L.

### Do the volume units need to match?

Yes, V&sub1; and V&sub2; must share a unit and C&sub1; and C&sub2; must share a unit. Millilitres with molarity is fine; millilitres against litres is not.

### Is C&sub1;V&sub1; = C&sub2;V&sub2; always valid?

It holds whenever the solute is conserved and volumes add. It drifts by about 4% for ethanol–water mixtures and more for concentrated acids.

### How much water do I add to 100 mL of 1 M stock to get 0.1 M?

About 900 mL, bringing the total to 1,000 mL. Adding a full litre gives 1,100 mL at 0.0909 M instead.

### What does a 1:10 dilution mean?

One part stock plus nine parts diluent, so ten parts total and a tenfold reduction. 10 mL of 1 M stock in 100 mL final volume gives 0.1 M.

### Can I dilute by adding solvent to the stock in the same flask?

Yes for a rough preparation, but a volumetric flask is better. Adding 900 mL to 100 mL assumes the volumes add exactly, which introduces up to 1% error with organic solvents.

### How do I get from 1 M to 1 mM?

Three tenfold steps: 1 M to 0.1 M to 0.01 M to 0.001 M. Alternatively take 1 mL of stock to 1,000 mL in one step, which is harder to measure accurately.

### Why is my diluted solution the wrong concentration?

Usually the final volume was treated as the solvent added. Topping up to 1,000 mL is not the same as adding 1,000 mL, and the gap is the 100 mL of stock.

### Does dilution change the number of moles?

No. 100 mmol of solute stays 100 mmol whether it sits in 100 mL or 1 L. Only the concentration, 1 M versus 0.1 M, changes.

## Method and sources

Methodology: Physical constants follow CODATA 2022 recommended values, and formulas follow standard physics definitions.

Last updated: 2026-11-17.
- Dilution equation (C₁V₁ = C₂V₂) for constant solute amount.